Add To Cart
share
Sick TCT-1AAT41006MZ Overview
TEMPERATURE SENSOR
Sick TCT-1AAT41006MZ Description
- Pt100 element, accuracy class A according to IEC 60751
- Measuring ranges -50 C ... 150 C and -50 C 250 C
- Wetted parts made from corrosion resistant stainless steel 316L
- Integrated transmitter with output signal 4 20 mA (2-wire)
- Measurement accuracy <= 0.2 % of span
- Various mechanical adaptations and insertion lengths, also available with thermowell
- Circular connector M12x1 (IP 67) or L-connector according to DIN EN 175301-803 A (IP 65)
- Reliable operation through rugged design and high-quality materials
- Good long-term stability, accuracy and linearity
- Quick and safe installation
- Convenient system integration through compact dimensions and industry-standard output signals
- Optimal solutions for individual requirements
Your benefits
| Features | |
| Measurement range: | -50 C ... 150 C |
| Sensor element: | Pt100 |
| Signal output and maximum ohmic load RA: | 4 mA ... 20 mA, 2-wire (RA <= (L+ - 9 V) / 0.023 A [Ohm]) |
| Performance | |
| Accuracy of sensor element: | class A according to IEC 60751 |
| Transmitter accuracy: | ≤ 0.2 % of span |
| Transmitter linearity: | ≤ 0.1 % of span |
| Mechanics/electronics | |
| Process connection: | thermowell 1/4'' NPT |
| Insertion length/diameter of probe: | 100 mm / 6 mm |
| Housing material: | Stainless steel 1.4571 |
| Wetted parts: | Stainless steel 1.4571 |
| Electrical connection/enclosure rating 1): | plug M12 x 1, 4-pin, IP 67 |
| Measuring current: | 0.3 mA ... 1 mA |
| Electrical safety: | protection class: III, dielectric strength: 500 V, reverse polarity protection: L+ towards M |
| Weight: | ca. 145 g (depending on configuration) |
| Supply voltage: | 10 V DC ... 36 V DC for variant with transmitter 4 mA ... 20 mA |
| Protection class III: | ![]() |
| CE-conformity: | 2004/108/EC, EN 61326-2-3 |
| Pressure resistance: | max. 220 bar |
| 1) Enclosure rating IP per IEC 60529. The enclosure rating classes specified only apply while the pressure transmitter is connected with female connectors that provide the corresponding enclosure rating. | |
| Ambient data | |
| Ambient temperature: | -40 C ... 85 C |
| Storage and transport temperature: | -40 C ... 85 C |
| Shock resistance according to IEC 60751: | 500 g |
| Vibration resistance according to IEC 60751: | 3 g |
Contact Us